Model DF-40L-400-Li
Emergency Response Digital Flowmeter, Air Sampler
Features:
- State-of-the-Art Electronics
- Operating Modes — Line Power (100VAC – 250 VAC) — On-Board Batteries
- Long-lasting 20 Ah Lithium ion battery pack
- Internal battery charging system operable from line power input
- Lightweight — ~16.5 lbs. (7.4 kg)
- Maximum Flow 30 to 35 LPM Maximum
- Bright LED Display
- Automatic Flow Control Feature
- Flowrate and Volume totalizations displayed are corrected to a factory settable Reference
- Temperature and Pressure (4 options available)
- Auto shut-off on time or volume
- Battery Life Indicator
F&J Model DF-40L-400-Li is a lightweight, small footprint, DC voltage powered air sampling system operable from (1) on-board 20 Ah Li Ion battery, (2) line power, or (3) auxiliary DC connector (12 – 15 VDC) The DF-40L-400-Li is well suited for emergency response sampling activities where users do not know whether line power will be available. Typical operating time from on-board battery is approximately 18 - 20 hours utilizing FP47M glass fiber filter paper. The automatic flow control feature with the F&J Digital Flowmeter technology provides maximum flexibility for emergency response sampling responsibilities. The set flowrate is maintained automatically in case of dust loading and does not require operator attention. The air sampler is contained in a lightweight small dimensioned enclosure which is water tight. The unit can be carried very easily by its large, rugged plastic handle. Typical flow range is 10 - 35 SLPM (0.3 -1.24 SCFM).
Pump Type: Dual diaphragm high efficiency
Maximum Flowrate: 35 LPM typical
with FP-47M glass fiber media
Power Source:
Internal – Li Ion; 14.8 VDC/20 Ah Battery Pack
External – Line Power; 100VAC to 240VAC
Battery Charging System:
Internal system which charges from line
power.
Operating Time on Batteries: ~ 18-20 hours
Current Draw: 0.6 A maximum
Filter Holder Fitting: 3/8 FNPT quick disconnect
Handle: Durable plastic
Weight: 16.5 lbs. (7.4 kg)
Dimensions: 16.8 in L × 9.0 in W × 14.0 in H (42.7 × 22.9 × 35.6 cm)
Average Db: 67.5
Operating Temperature Range: 0ºF to 122ºF (-18ºC to 50ºC)
Display: Bright LED (6 character, 1,2 cm H)
Elapsed Time: DD:HH:MM up to 168 hours
Flow Control:
- Adjustable from keypad between 10-35 LPM
- Flow Accuracy: ±4.0% of Full Scale
- Factory Settable Reference T and P
- Classical STP 0ºC, 1 ATM
- Normal T and P 20ºC, 1 ATM
- Modified Normal T and P 70ºF, 1 ATM
- Standard Ambient T and P 25ºC, 1 ATM
- 2 GB Secure Digital Card (P/N: 372239)
- Flash card Reader (P/N: SDDR-199-A20)
- Combination Filter Holder with Rain Hood
- 47mm Particulate / C Geometry – FJ-47RH
- 50mm Particulate / C Geometry – FJ-50RH
Filter Holder Model | Charcoal Cartridge Dimensions | Particulate Paper Diameter |
---|---|---|
FJ-05P | F&J Model B | 2 in or 50 mm |
FJ-21P | F&J Model C | 2 in or 50 mm |
FJ-35P | F&J Model B | 47 mm |
FJ-46P | F&J Model C | 47 mm |
FJ-51P | F&J Model M | 2 in or 50 mm |
FJ-53P | F&J Model M | 47 mm |